• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DITRON DENTAL LTD ULTIMATE PRECISION IMPLANT; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DITRON DENTAL LTD ULTIMATE PRECISION IMPLANT; DENTAL IMPLANT Back to Search Results
Model Number ULT
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Circumstances are not clear due to the lack of provided information.To date, no information has indicated that the device or the way it was used has contributed to the described event.In addition, no signals were found indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to osseointegration problem, and escaping the available detections has been assessed and concluded to be unlikely in high level of confidence.If additional information is received indicating otherwise, a follow-up report will be submitted.
 
Event or Problem Description
Loss of osseointegration, instability of the implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTIMATE PRECISION IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
DITRON DENTAL LTD
2 haofe st.
southern industrial zone
ashkelon, israel 78780 02
IS  7878002
Manufacturer (Section G)
DITRON DENTAL LTD
2 haofe st.
southern industrial zone
ashkelon, israel 78780 02
IS   7878002
Manufacturer Contact
2 haofe st.
southern industrial zone
ashkelon, israel 78780-02
MDR Report Key20804170
Report Number3010948854-2022-00270
Device Sequence Number21145856
Product Code DZE
UDI-Device Identifier07290109280639
UDI-Public07290109280639
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K161497
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/06/2026
Device Model NumberULT
Device Catalogue Number11375
Device Lot NumberULT000240
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2022
Initial Report FDA Received Date11/28/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-