| Catalog Number |
TSVMB8 |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).A1: patient identifier: unknown / not provided.A2: age at time of the event: unknown / not provided.A3: gender: unknown / not provided.A4: patient weight: unknown / not provided.B3: date of event: unknown / not provided.D10: concomitant medical product: tsx37b11 (x10), tsx implant, 3.7mmd, 11.5mml (x10), lot: 1275485 (x10) tsvmb11, imp,tsv,mcol mg,3.7mm,11.5mml, lot: 1270239.D10: therapy date: unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that the dentist opened 3 new product packages in front of the territory manager they were all 3 with the inner tub cap broken.A total of 12 implants with the inner tube cap broken were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number cmp-(b)(4).The following fields have been updated: a: patient identifier b4: date of this report b5: describe event or problem g3: date received by manufacturer g6: type of report h1: type of reportable event h2: follow up type h3: device evaluated by manufacturer h6: adverse event problem h10: additional narrative zimvie received one (1) tsvmb8, (imp,tsv,mcol mg,3.7mm,8mm) for evaluation.A visual evaluation was performed, the reported implants packaging matched the report.The implants outer vials tamper seals were not broken.The outer vials green cap(s) were fractured.The coob is confirmed.Device history record (dhr) review was completed for the subject lot number 1267722.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1267722 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : packaging : damaged.The customer did submit images for the reported event.The images showed the implant outer vials green cap was cracked.Based on the investigation and according to the capa, the most likely root cause determined from the investigation was external factors (temperature).This is an ongoing issue which is currently being monitored.Non conformance and health hazard evaluation have been opened to further contain affected products, evaluate potential root causes, and consider applicable corrective actions.Therefore, based on the available information, a packaging malfunction did occur.The implants cap was cracked while the caps tamper seal was intact.The reported event/coob was confirmed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|