MEDTRONIC XOMED INC. XPS® HANDPIECE - SKEETER® OTOLOGIC DRILL; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I Back to Search Results
Model Number 3055601
Medical Device Problem Code Vibration (1674)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/05/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that device was returned for planned maintenance.There was no patient involved.
 
Additional Manufacturer Narrative
Analysis found that skeeter handpiece vibrates and runs rough.The motor assembly was faulty.The cable was worn.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® HANDPIECE - SKEETER® OTOLOGIC DRILL
Common Device Name
DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
justin ellis
6743 southpoint drive north
jacksonville, FL 32216
7635265677
MDR Report Key20808474
Report Number1045254-2024-02070
Device Sequence Number13219224
Product Code ERL
UDI-Device Identifier00643169882676
UDI-Public00643169882676
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K833874
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3055601
Device Catalogue Number3055601
Device Lot Number219715361
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/05/2024
Initial Report FDA Received Date11/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/16/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-