• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS POWERLED TM; LIGHT, SURGICAL, CEILING MOUNTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS POWERLED TM; LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Model Number ARD568410710A
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.
 
Event or Problem Description
Getinge became aware of an issue with one of surgical lights - powerled.As it was stated, the paint issue was found on the suspension arm and some of the paint particles fall down.It was established by technician that probably wrong cleaning materials were used.The designated complaint unit employee confirmed the issue based on photographic evidence.There was no injury reported, however, we decided to report the issue in abundance of caution as any particles falling off into sterile field or during procedure may cause contamination in case of reoccurrence.
 
Additional Manufacturer Narrative
Getinge became aware of an issue with one of surgical lights - powerled.As it was stated, the paint issue was found on the suspension arm and some of the paint particles fall down.It was established by technician that probably wrong cleaning materials were used.The designated complaint unit employee confirmed the issue based on photographic evidence.There was no injury reported, however, we decided to report the issue in abundance of caution as any particles falling off into sterile field or during procedure may cause contamination in case of reoccurrence.Based on the information collected, it was established that when the event occurred, the surgical light did not meet its specification and in this way the device contributed to event.It is unknown if claimed devices were or were not being used for patient treatment or diagnosis when the event took place.According to subject matter expert at manufacturer¿s, the most probable root cause of this premature paint degradation would be an incorrect cleaning procedure/product.All maquet sas products comply with: ¿ iec 60601-1 ed.2.0 & ed.3.0 general requirements for basic safety and essential performance.¿ iec 60601-2-41 particular requirements for the safety of surgical luminaires and luminaire for diagnosis.¿ paint definition pfc066.This procedure defines maquet sas's requirements for all painted parts.¿ disinfection products test: the aim of these tests is to detect any incompatibility with disinfectant.This event is clearly due to the cleaning product.Nevertheless, the deterioration of the paint is far advanced, and it must have started years ago.Such troubles should have been detected first by the user during daily and monthly checks.The current cleaning product must be improved as described in the user manual.The damaged parts must be replaced as soon as possible.Some of the paint chip or paint damages are due to: impacts, collisions (abnormal use).The operating manual includes the instructions to pre-position the arms prior to use, in order to prevent damages.To prevent any similar incident, it is recommended to avoid collisions between devices.Visual inspections during the cleaning allow to detect the painting defect, we recommend to perform corrective maintenance to rectify the default after its detection.Minor paint chip can be repaired with touch up paint, nevertheless, the parts impacted by serious damage must be replaced.Getinge shall continue to monitor for any further events of this nature and does not propose any further action at this time.
 
Event or Problem Description
Manufacturer's reference number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERLED TM
Common Device Name
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
pascal jay
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key20808909
Report Number9710055-2024-0000607
Device Sequence Number18625700
Product Code FSY
UDI-Device Identifier03700712402969
UDI-Public(01)03700712402969
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K070442
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberARD568410710A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/26/2024
Supplement Date Received by Manufacturer04/10/2025
Initial Report FDA Received Date11/29/2024
Supplement Report FDA Received Date04/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/22/2017
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-