| Model Number |
400-0100.07 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/30/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This case was reviewed and investigated according to the manufacture¿s policy.Blocks e1 & g2: country is china.Block h3: at the customer site, a field service engineer replaced the burnt fuse and confirmed the system is working normally.Block h6: during use of the core mobile, the system stopped working due to a component failure without any design or manufacturing issue.Do not apply to this submission.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
| |
|
Event or Problem Description
|
|
It was reported that a core mobile system was used in a coronary diagnostic emergent procedure.During preparation for use, the equipment fuse burned out and the system shut down.The case was completed without the use of ivus.No injury to patient or user was reported.This product problem is being reported because the system could not be used during an emergent case due to a burnt fuse; resulting in a potential for harm.
|
| |
|