Model Number 6900PTFX
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code Dyspnea (1816)
Date of Event 10/24/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was learned through implant patient registry and investigation that a patient with a 31mm 6900ptfx mitral valve underwent a valve-in-valve procedure after an implant duration of five (5) years, six (6) months due to degeneration and stenosis.Patient presented with shortness of breath.The procedure was performed with a 29mm 9755rsl transcatheter valve.Patient was stable, in recovery and was discharged on pod#1.This is a 79-year-old male patient edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional 'customer complaint'.The information reported may or may not be related to the edwards device.
 
Additional Manufacturer Narrative
H11.Additional narrative: updated d4, h4, and h6.The device history record (dhr) was reviewed and showed that this device met all manufacturing and sterilization specifications for product release prior to distribution.No issues were identified that would have impacted this event.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES, LLC
one edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES, LLC
one edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2 - office m2013
irvine, CA 92614
9492506615
MDR Report Key20815772
Report Number2015691-2024-09056
Device Sequence Number18063139
Product Code DYE
UDI-Device Identifier00690103163612
UDI-Public(01)00690103163612(17)201214
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/14/2020
Device Model Number6900PTFX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/05/2024
Supplement Date Received by Manufacturer12/04/2024
Initial Report FDA Received Date12/01/2024
Supplement Report FDA Received Date12/31/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/15/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention; Hospitalization;
Patient Age79 YR
Patient SexFemale
-
-