| Model Number |
REF 316-07 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
According to the complaint description it was noticed after 1 day of use that the subglottic port was split.A theoretical investigation was conducted, as no photo/sample and no further information was available to date.No verification of defect possible to date.The production data is inconspicuous.The port can be damaged if the syringe is inserted too tightly.However, this might be also an injection molding defect, which is very rare.Further investigation within capa.
|
| |
|
Event or Problem Description
|
|
After 1 days of use the port of the suction line of the tracoe twist plus tracheostomy cannula split.Difficult to obtain aspirates so planned elective change performed the next day.No perceived harm to patient as a result.
|
| |
|