Based on our review of the device history record (dhr) for the subject device, there were no recorded nonconformances, deviations, or other manufacturing anomalies that could account for the reported failure.All production and inspection processes were completed in accordance with specified requirements, and no irregularities were identified that would have impacted the device's functionality or quality.Sterile samples from the reported lot, including the actual devices, were not returned for visual review or analysis.Without the actual reported devices, photos of the devices, or detailed information on the method used to remove the device from the packaging, preoperative preparation of the device, or the surgeon¿s technique, a definitive root cause cannot be determined at this time.Review of labeling 03-5278r10: contraindications: quill¿ knotless tissue-closure device comprised of dyed pdo is not to be used where prolonged (beyond six weeks) approximation of tissues under stress is required and is not to be used in conjunction with or for fixation of prosthetic devices (e.G.Heart valves or synthetic grafts) that are non-absorbable in nature.Precautions: quill¿ knotless tissue-closure device contains bidirectionally oriented barbs to anchor tissues and does not require knots to approximate opposing edges of a wound.Tying knots on the barbed section of the material will damage the barbs and potentially reduce the suture tensile strength and barb effectiveness.For the bidirectional forces to be created and for the device to function properly, both sides of the quill¿ knotless tissue-closure device must be engaged in the tissue.Additionally, when completing placement, an additional backstitch or bite of tissue lateral to the end of the incision is required to lock the device in place.Avoid contacting the quill¿ knotless tissue-closure device and associated needles with other materials (e.G.Surgical gauze, drapes, etc.) in the surgical field to prevent ensnaring on the barbs.If the barbs catch, carefully pull the material in the opposite direction of the needle to disengage it from the barbs.Care should be taken to avoid damage when handling.Avoid crushing or crimping the suture material with surgical instruments, such as needle holders and forceps.Do not pull the quill¿ knotless tissue-closure device comprised of dyed pdo out of the package by the needles as this can cause the barbs to catch on one another.Do not attempt to remove memory in the polymer by running fingers down the suture material as this can damage the barbs.Infections, erythema, foreign body reactions, transient inflammatory reactions and in rare instances wound dehiscence are typical or foreseeable risks associated with any suture and hence are also potential complications associated with quill¿ knotless tissue-closure device comprised of dyed pdo.When using quill¿ knotless tissue-closure device comprised of dyed pdo subcutaneously, the device should be placed as deeply as possible in order to minimize erythema and induration normally associated with absorption.Acceptable surgical practice should be followed with respect to drainage and closure of infected wounds.To avoid damaging needle points and swage areas, grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the swaged end to the point.Reshaping needles may cause them to lose strength and be less resistant to bending and breaking.Users should exercise caution when handling surgical needles to avoid inadvertent needle sticks.Discard used needles in ¿sharps¿ containers.Adverse reactions: adverse effects associated with the use of this device may include, wound dehiscence, failure to provide adequate wound support in closure of the site where expansion, stretching or distension occur, failure to provide adequate wound support in elderly, malnourished or debilitated patients or in patients suffering from conditions which may delay wound healing, infection, minimal acute inflammatory tissue reaction, localized irritation when skin sutures are left in place for greater than 7 days, suture extrusion and delayed absorption in tissue with poor blood supply, calculi formation in urinary and biliary tracts when prolonged contact with salt solutions such as urine and bile occurs and transitory local infection as the wound site.Broken needles may result in extended or additional surgeries or residual foreign bodies.Inadvertent needle sticks with contaminated surgical needles may result in the transmission of bloodborne pathogens.Due to prolonged suture absorption, some irritation and bleeding may occur.
|