Additional information is provided in sections d.9, h.3, h.6 and h.11.Six unopened 2.75mm ophthalmic slit knives, in blisters were received for the report of dull blades.Samples were visually inspected and was found to be conforming.Functional penetration testing was then performed and all were found to be conforming.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The returned unopened samples were found to be conforming, therefore dull blades was not confirmed.However since the actual complaint samples were not returned, a root cause cannot be determined for the complaint as described by the customer.The unopened samples were found conforming; however because the actual complaint samples were not returned the exact root cause for this complaint is unknown, specific action with regards to this complaint cannot be taken.All knives are 100% inspected by trained operators using a minimum of 10x magnification during manufacturing.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|