Model Number 2H7462
Medical Device Problem Code Backflow (1064)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a clearlink system non-dehp secondary medication set backfilled into a clearlink system continu-flo primary solution set.The piggyback secondary set was connected to the primary set during infusion when it was observed that the infusion was not going to the patient but backfilled into the primary tubing chamber.To resolve this issue, ¿hooked the tubing up distal to the roller clamp and turned the roller clamp off to the primary solution to allow the medication to infuse¿.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
E1: initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
Additional information: h4, h6, h11.H11: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK/DUO-VENT
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
600 mts oeste de entrada
las americas parque industrial
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key20819729
Report Number1416980-2024-06913
Device Sequence Number20742381
Product Code FPA
UDI-Device Identifier00085412048543
UDI-Public(01)00085412048543(10)R24H22086(17)260823
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K203609
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 12/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/23/2026
Device Model Number2H7462
Device Catalogue Number2H7462
Device Lot NumberR24H22086
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/04/2024
Supplement Date Received by Manufacturer12/15/2024
Initial Report FDA Received Date12/02/2024
Supplement Report FDA Received Date12/30/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-