| Model Number |
MAF-TM2 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1 (facility name): (b)(6) hospital.E2/e3: no information available for the occupation of the initial reporter or whether they are a healthcare professional.The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the endotracheal tube became stuck on the subject device during a lab demonstration.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Updated fields: d8, h2, h4, h6, h11.The device was returned to olympus for inspection and the reported failure was confirmed.Based on the results of the investigation, a definitive root cause could not be identified.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No new additional information received from the customer.
|
| |
|