Baxter received a report from a baxter technician to report the compella air mattress, na, r power cord sparked and smoked due to short.The bed was located at the account.This occurred during patient use.The reporter did not know if there was any patient injury or medical intervention associated with this event.It was unknown if any second device was also involved.The customer did not know if this event was already communicated to another baxter department or authority.
|
|
The baxter technician found the power cord needed to be replaced.Per the bater service manual it is necessary for the compella bariatric bed to have an effective maintenance program.We recommend that you do annual preventive maintenance.Pay particular attention to safety features, which include but are not limited to: the power cords for the bed and asu are with the unit, and they are in good condition: the plug is a one-piece molded plug assembly, the assembly shows no signs of cracking, the plug molding around the blade is not discolored, and the blade is tight in the molding.The power cord hooks shows no sign of cracking and are intact with no damage.Replace or repair parts as necessary.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the power cord to resolve the reported event.Based on this information, no further action is required.
|