MPRI LEAD END CAP KIT; PACEMAKER LEAD ADAPTOR Back to Search Results
Model Number 5867-3M
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Erosion (1750); Unspecified Infection (1930)
Date of Event 10/22/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the previously capped pace/sense portion of the right ventricular (rv) lead eroded through the skin and was the source of infection.The lead and lead cap were explanted and replaced with a previously capped rv pace/sense lead plugged into the pace/sense port of the cardiac resynchronization therapy defibrillator (crt-d).Detections were deactivated, the high voltage port was plugged, and the system remains in use for pacing support.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10: 5054-52 lead; implanted: (b)(6) 2001, 5554-45 lead; implanted: (b)(6) 2001, dtpa2q1 crt-d; implanted (b)(6) 2023.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD END CAP KIT
Common Device Name
PACEMAKER LEAD ADAPTOR
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key20821406
Report Number2649622-2024-32415
Device Sequence Number8955378
Product Code DTD
UDI-Device Identifier00763000059460
UDI-Public00763000059460
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K911048
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date08/25/2025
Device Model Number5867-3M
Device Catalogue Number5867-3M
Device Lot NumberVA2GBWL
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/11/2024
Initial Report FDA Received Date12/02/2024
Date Device Manufactured08/31/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
693558 LEAD, 429878 LEAD
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age76 YR
Patient SexMale
-
-