ESTABLISHMENT LABS S.A MOTIVA BREAST IMPLANT ROUND FULL WITH QID ERGONOMIX 275 S; MOTIVA IMPLANTS Back to Search Results
Catalog Number ERSF-275Q
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Rupture (2208); Insufficient Information (4580)
Date of Event 10/31/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported a female patient who underwent breast implantation surgery with a motiva implant experienced left-sided rupture according to mri surveillance.No patient demographics, such as age, weight, or ethnicity, were provided by the reporter, and no adverse outcomes for the patient have been noted.Efforts to gather additional information are ongoing, and the investigation remains in progress.
 
Additional Manufacturer Narrative
As stated in the literature, ¿a number of risk factors for rupture have been identified; the most common cause is surgical instrument damage.¿ (handel et al., 2013).Additionally, rupture is a common and known reason for complaints, and it is clearly characterized in the product dfu included with the implant, as stated above.Additionally, no cases of rupture because of product failure have been reported to establishment labs ever.Per our directions for use, each patient must receive the establishment labs s.A.¿motiva implants®: information for the patient¿ during her surgical consultation, it is the surgeons´ responsibility to ensure that the patient completely understands the information regarding the risks, benefits, and recommendations associated with silicone gel-filled breast implant surgery, as well as the complications typical of any type of surgery.This document is available in motiva® website: ifu.Motiva.Health the instructions for use in the directions for use were reviewed, to determinate if there are indications that ensure the prevention and good handling of the product.For this event, it is considered that the information is suitable for the section of surgical precautions as follows: ¿breast implants can potentially remain intact for decades in the body, but all such devices will fail at some point.Breast implants rupture when the shell develops a tear or hole.Rupture can occur at any time after implantation, but it is most likely to occur; the longer the implant is in place.The following may cause implants to rupture: damage by surgical instruments, implant stress, and weakening during implantation, age and design of the implant, submuscular rather than sub-glandular location, the occurrence of post-operatory hematomas or seromas, folding or wrinkling of the implant shell, excessive force to the chest, trauma, compression during mammographic imaging, and severe capsular contracture¿.Establishment labs requested the unit to confirm the event and to perform the corresponding testing (including but not limited to: revision and characterization under the microscope of the rupture section, mechanical testing according to approved standards, etc.).However, it was informed that the patient hasn´t been explanted.Breast implants are not considered lifetime devices, and implant rupture is a risk associated with breast surgery, which can occur at any time.
 
Additional Manufacturer Narrative
While the product was not returned for evaluation, there was a relationship between the reported event and the device based on review of the evidence provided.A clinical evaluation of the report and evidence was executed and a rupture was confirmed.A complete review of the dhr for lot 20111129 was carried out, no deviation was found in the manufacturing process.Review of complaint history for the reported lot number and sterilization run found no other similar complaints reported in the past.¿ while the rupture was confirmed, a definitive root cause could not be determined.Potential factors non-related to the manufacture of the device that may lead to the events reported include, but are not limited to: (1) damage by surgical instruments.(2) implant stress and weakening during implantation.¿ as no manufacturing, material or design issues were identified, no further investigation or additional actions are required at this time.¿ establishment labs will continue to monitor for similar complaints/trends.
 
Event or Problem Description
Mexico.It was reported a female patient who underwent breast implantation surgery with a motiva implant experienced left-sided rupture according to mri surveillance.No patient demographics, such as age, weight, or ethnicity, were provided by the reporter, and no adverse outcomes for the patient have been noted.Efforts to gather additional information are ongoing, and the investigation remains in progress.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTIVA BREAST IMPLANT ROUND FULL WITH QID ERGONOMIX 275 S
Common Device Name
MOTIVA IMPLANTS
Manufacturer (Section D)
ESTABLISHMENT LABS S.A
coyol free zone, building b25
alajuela CA 93101
Manufacturer (Section G)
ESTABLISHEMENT LABS
building b-25, coyol free zone
alajuela
Manufacturer Contact
susana hidalgo
coyol free zone, building b25
alajuela, CA 93101
8003554164
MDR Report Key20825407
Report Number3012883202-2024-00169
Device Sequence Number7813294
Product Code FTR
UDI-Device Identifier07445161012326
UDI-Public(01)07445161012326(11)201209(17)251209
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/09/2025
Device Catalogue NumberERSF-275Q
Device Lot Number20111129
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2024
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date12/02/2024
Supplement Report FDA Received Date03/03/2025
10/27/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/09/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-