• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON SYRINGE 5ML HEPARIN 10UNIT; HEPARIN, VASCULAR ACCESS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON SYRINGE 5ML HEPARIN 10UNIT; HEPARIN, VASCULAR ACCESS FLUSH Back to Search Results
Catalog Number 306414
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2024
Type of Reportable Event Malfunction
Event or Problem Description
During locking the line nursing tried to use posi flush heparin 5ml (10units) but it was observed stopper getting detached from plunger.
 
Additional Manufacturer Narrative
Initial mdr submission.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
 
Additional Manufacturer Narrative
(b)(4) follow up.A device history record review was completed by our quality engineer team for provided material number 306414 and lot number 415861n.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.The product is designed to push the plunger rod down to expel the solution.It is not for pulling the plunger rod back.Should you again experience any problems with our product we would appreciate the opportunity to conduct a thorough analysis.There are quality controls currently in place to detect this type of defect during the production process.Further action has not been determined necessary at this time.
 
Event or Problem Description
No additional information received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 5ML HEPARIN 10UNIT
Common Device Name
HEPARIN, VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key20827874
Report Number1911916-2024-00863
Device Sequence Number8502231
Product Code NZW
UDI-Device Identifier30382903064145
UDI-Public(01)30382903064145(11)240606(17)251130(10)415861N
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K090680
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2025
Device Catalogue Number306414
Device Lot Number415861N
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/25/2024
Supplement Date Received by Manufacturer12/23/2024
Initial Report FDA Received Date12/02/2024
Supplement Report FDA Received Date12/28/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/06/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-