• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS 1-PIECE IOL WITH TECNIS SIMPLICITYTM DELIVERY SYSTEM; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS 1-PIECE IOL WITH TECNIS SIMPLICITYTM DELIVERY SYSTEM; INTRAOCULAR LENS Back to Search Results
Model Number DCB00
Medical Device Problem Codes Break (1069); Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Section a5: per regulation eu 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.Section b3 - date of event: unknown, as information was requested but not provided.Section d6a - implant date: n/a - lens was not implanted.Section d6b - explant date: n/a - lens was not implanted.Section e1 - telephone number: (b)(6).Device evaluation: product testing could not be performed since the device was not returned for evaluation (lens was discarded at the customer site).Therefore, the reported issue could not be verified, and product quality deficiency could not be determined.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.A search of complaints related to this production order (po) was performed.The search revealed that no additional complaints were received for this po.No escalation required.Conclusion: a product deficiency has not been identified; therefore, no additional corrective actions have been initiated.Attempts have been made to obtain the missing information.However, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the preloaded, monofocal, intraocular lens (iol) would not advance.Through follow up we learned that during preparation of the device, the surgeon noted the iol to be crumpled inside the cartridge and decided to discard.The procedure was completed using a back up iol.There was no patient contact reported.No further information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1-PIECE IOL WITH TECNIS SIMPLICITYTM DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key20827880
Report Number3012236936-2024-000299
Device Sequence Number18268227
Product Code HQL
UDI-Device Identifier05050474635913
UDI-Public(01)05050474635913(17)260622
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 12/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/22/2026
Device Model NumberDCB00
Device Catalogue NumberDCB0000115
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2024
Initial Report FDA Received Date12/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/22/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-