Catalog Number 8065752216
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/17/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that bent tip was observed during cataract (with intraocular lens implant) surgery.The surgery was performed and completed after replacing the product with another one.Patient impact was not reported.
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.11.One opened phaco tip, without a wrench, in a small parts tray, in a tray was received.Sample was visually inspected and found to be conforming.No nonconforming bent condition was observed.Wear observed on threads, back of flange, and nut corners consistent with threading on handpiece.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The complaint evaluation does not confirm the report of the bent tip during surgery.The returned sample was found to be visually conforming, therefore a bend phaco tip as described in the complaint was not confirmed and a root cause cannot be determined for the complaint as described by the customer.No action was taken as the phaco tip was manufactured to specification.All phaco tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key20829378
Report Number1644019-2024-02898
Device Sequence Number4032366
Product Code LRO
UDI-Device Identifier00380657522163
UDI-Public(01)00380657522163(17)251031(11)231130(10)16AJHM
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Catalogue Number8065752216
Device Lot Number16AJHM
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer11/26/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/05/2024
Supplement Date Received by Manufacturer02/03/2025
Initial Report FDA Received Date12/03/2024
Supplement Report FDA Received Date02/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/30/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-