• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA USA INC. CARDIOPLEGIA PACK; SMART, PERFUSION TUBING PACKS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA USA INC. CARDIOPLEGIA PACK; SMART, PERFUSION TUBING PACKS Back to Search Results
Catalog Number 044009103
Medical Device Problem Code Air/Gas in Device (4062)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/04/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Livanova usa inc.Received a report related to an air ingress into a perfusion tubing system (pts) pack.In detail, cardioplegia line was primed and flushed with no leaks noted when over-pressurized while in priming mode.Once cross clamp was on, while delivering the cardioplegia solution to the patient, the delivery system and line filled with air.Delivery was then stopped and attempted to flush air to the field.However, more air came into delivery system.Therefore, cross clamp was removed.Patient was defibrillated and cardioplegia system was exchanged for new one.New cardioplegia circuit was primed and flushed to the field.Cross clamp was placed and cardioplegia was delivered with no air issues.
 
Additional Manufacturer Narrative
A.1.-a.5.Patient information was not provided.H11: livanova usa inc.Manufactures the cardioplegia pack.The incident occurred in (b)(6).Through follow-up communication livanova learned that the affected unit is not available for investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Additional Manufacturer Narrative
H11: by reviewing the complaints database, no further similar events have been recorded on this perfusion tubing system (pts) catalogue number nor a concerning trend has been identified.Review of the device history record (dhr) for the affected lot did not reveal any deviation or non-conformity.Based on the available information and without the part to investigate the root cause of the event cannot be established.Considering that no issue occurred during priming, it is unlikely that a defective component or connection was initially present on the cardioplegia line, but a defect could have originated due to stresses and manipulation applied during procedure.
 
Event or Problem Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOPLEGIA PACK
Common Device Name
SMART, PERFUSION TUBING PACKS
Manufacturer (Section D)
LIVANOVA USA INC.
14401 w. 65th way
arvada CO 80004
Manufacturer (Section G)
LIVANOVA USA INC.
14401 w. 65th way
arvada CO 80004
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key20830131
Report Number1718850-2024-00045
Device Sequence Number14461141
Product Code DWE
UDI-Device Identifier00803622142550
UDI-Public(01)00803622142550(240)044009103(17)260227(10)2405900038(11)240228
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K981613
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 02/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/27/2026
Device Catalogue Number044009103
Device Lot Number2405900038
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/02/2024
Supplement Date Received by Manufacturer01/31/2025
Initial Report FDA Received Date12/03/2024
Supplement Report FDA Received Date02/27/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/28/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-