• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE POS FEM AUG SZ 8 4MM; KNEE FEMORAL ACCESSORY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE POS FEM AUG SZ 8 4MM; KNEE FEMORAL ACCESSORY Back to Search Results
Catalog Number 154908001
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hypersensitivity/Allergic reaction (1907)
Date of Event 11/27/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #(b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that as per surgeon, the patient had nickel allergy.Surgeon removed all components and put in competitor's nickel free implants.There was no surgical delay.Doi: (b)(6) 2022.Dor: (b)(6) 2024.Affected side: right knee.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Added: d10 concomitant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE POS FEM AUG SZ 8 4MM
Common Device Name
KNEE FEMORAL ACCESSORY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key20831298
Report Number1818910-2024-24709
Device Sequence Number14493027
Product Code JWH
UDI-Device Identifier10603295383123
UDI-Public(01)10603295383123(17)311031(10)JM8541
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number154908001
Device Lot NumberJM8541
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/11/2024
Supplement Date Received by Manufacturer12/04/2024
Initial Report FDA Received Date12/03/2024
Supplement Report FDA Received Date12/05/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CEMENTED STEM 14X50MM; ATTUNE CRS FEMORAL RT SZ 8 CEM; ATTUNE CRS RP INSRT SZ 8 6MM; ATTUNE POS FEM AUG SZ 8 4MM; ATUN TIB SLV M/L 37MM FULL POR; ATUNE PRESSFIT STR STEM18X60MM; ATUNE REV RP TIB BASE SZ 7 CEM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient SexMale
-
-