Catalog Number PLC121000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Rupture (2208)
Date of Event 11/06/2024
Type of Reportable Event Serious Injury
Event or Problem Description
The following was reported to gore: on (b)(6) 2024, the patient underwent an endovascular procedure to treat an aneurysm utilizing the gore® excluder® aaa endoprostheses and gore® molding and occlusion balloon catheter (mob) in zone 9.It was reported that the patient had calcified iliacs, especially on the left side.After placing the trunk ipsilateral component (rlt311413) up the right side and the contralateral leg component (plc121000) up left, the physician ballooned with a mob37 at the proximal neck and then at the left common iliac.The stent ended right above the iliac bifurcation.The proximal common was stenotic.It appears the mob inflation tore the iliac bifurcation.The stents were undamaged.Shortly after ballooning, the patient's pressure and heart rate dropped.They did compressions and transfusions along with epinephrine.They placed an 1159 gore® viabahn® vbx balloon expandable endoprosthesis but it wasn't placed high enough.The physician placed an iliac extender (pll16207) and ballooned with a mob37.This sealed the leak.The patient was stabilized and transferred to the intensive care unit (icu).The physician believes ballooning aggressively caused the reported issue.
 
Additional Manufacturer Narrative
According to the gore® excluder® aaa endoprosthesis instructions for use, adverse events that may occur and / or require intervention include vascular trauma (e.G., ilio-femoral vessel dissection, bleeding, rupture).W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL SILICON VALLEY B/P
2890 de la cruz blvd.
santa clara CA 95050
Manufacturer Contact
monique nwanze
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key20831485
Report Number3013164176-2024-02286
Device Sequence Number12944159
Product Code MIH
UDI-Device Identifier00733132618613
UDI-Public00733132618613
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberPLC121000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/06/2024
Initial Report FDA Received Date12/03/2024
Date Device Manufactured08/05/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexMale
-
-