Model Number 72018201
Medical Device Problem Codes Mechanical Problem (1384); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/25/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this inflatable penile prosthesis (ipp) was not performing as expected.Two weeks later a revision surgery was performed, upon examination a hole in the reservoir was found.The pump was removed and replaced, while the existing reservoir was kept in the patient and a new one was added to the system.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONCEAL
Common Device Name
PROSTHESIS, PENIS, INFLATABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key20831867
Report Number2124215-2024-75390
Device Sequence Number4032452
Product Code JCW
UDI-Device Identifier00878953005676
UDI-Public00878953005676
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/09/2022
Device Model Number72018201
Device Catalogue Number72018201
Device Lot Number0190538018
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/11/2024
Initial Report FDA Received Date12/03/2024
Date Device Manufactured10/23/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LOT 1000002270 UPN 72404273 CX.; LOT 185816002 UPN 72404310 PUMP.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age65 YR
Patient SexMale
-
-