|
This case was reviewed and investigated according to the manufacturer¿s policy.Block a3b: no information available.Block c: not applicable for this device.Blocks d6 & d7: not applicable for this device.Blocks h3 & h6: the visions catheter was not returned to the manufacturer for evaluation; thus no returned product investigation was performed.Blocks h7, h9 & h10: do not apply to this submission.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
|
Blocks d9 & h3: the visions pv.018 catheter was returned for evaluation.Block h3: the visions pv.018 catheter was returned in two pieces.The distal section includes a portion of the distal tip; measuring approx.0.9 cm in length.The proximal section includes a portion of the distal tip, distal fillet, scanner body, expanded single lumen, shaft, luer, and connector; measuring approx.136.5 cm in length.The total length combined is 137.4 cm, which is within specification of 135-139 cm.Visual inspection found no other damage observed.Block h6: the probable cause of the separation was likely damaged during use/handling.Manipulation, strain, impact, and forces associated with use can affect the integrity of the device.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|