| Catalog Number |
8187501 |
| Medical Device Problem Code |
Restricted Flow rate (1248)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
While using a cavitron jet plus w/tapon-115v g137 they allege that they have inconsistent water flow and the handpiece is heating up, no injury resulted.
|
| |
|
Additional Manufacturer Narrative
|
|
12/18/2024 repair tech: (b)(4).000 evaluated, meets specifications; contact customer.Cleaned and retubed powder bowl.Jet mate nozzle was clogged and needed replaced.Pointer in powder bowl cap was replaced, cracked.Qa-mo.***hpc-10230*** ***jm-10230*** dhr review is not required because the product was returned for evaluation.The service technician could not replicate the failure noted in the complaint with the returned unit.No additional action is necessary.
|
| |
|