| Model Number |
N/A |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/08/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: cat# 010000896 lot# 6694887 g7 10 deg e1 liner 36mm e cat# 00489430000 lot# 64532856 trabecular metalâ¿¢ acetabular revision system, column buttress, left posterior / right anterior cat# 00489405010 lot# 64578576 50mm x 10mm thickness acetabular augment cat# 00625006520 lot# j6872908 bone screw self-tapping 6.5 mm dia.20 mm length cat# 00625006530 lot# j6915128 bone screw self-tapping 6.5 mm dia.30 mm length cat# 00662406520 lot# 62460740 bone screw self-tapping 20 mm length 6.5 mm dia.Cat# 00662406520 lot# 62641023 bone screw self-tapping 20 mm length 6.5 mm dia.Cat# 00662406520 lot# 64257747 bone screw self-tapping 20 mm length 6.5 mm dia.Cat# 00662406530 lot# 65270239 bone screw self-tapping 30 mm length 6.5 mm dia.Cat# 00662406530 lot# 65270239 bone screw self-tapping 30 mm length 6.5 mm dia.Cat# 00662406540 lot# 63704581 bone screw self-tapping 40 mm length 6.5 mm dia.G2: foreign: australia the customer has indicated that the product will not be returned to zimmer biomet for investigation as its location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an initial hip procedure.Subsequently, the patient was revised approximately 16 months later due to acetabular loosening.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected} visual examination of the provided pictures identified the explanted shell with bio-debris and what appears to be bone ingrowth on the surface.No further evaluation can be made.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: ct images of the left hip demonstrating anatomic alignment of the arthroplasty components but without solid bridging of the acetabular implant and cement noted consistent with implant loosening.In follow up, the surgeon noted space between acetabular component and host bone.A definitive root cause cannot be determined.This complaint was confirmed based on the mmi review confirming the loosening.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further information at the time of this report.
|
| |
|
Search Alerts/Recalls
|