|
It was reported that immediately after inserting the intra-aortic balloon (iab), at around 7:00 am on (b)(6) 2024, it was noticed that the iab did not expand.Even after changing the pump, the situation did not change.The iab was then replaced and pumping began normally and therapy was completed.There was no patient harm or adverse event reported.
|
|
Updated fields: b4, e2, e3, g1, g3, g6,g7, h2, h6 (type of investigation), h11 no decon or visual inspection performed since product was not returned and could not be evaluated.Therefore, we were unable to confirm or determine a root cause for the reported failure.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.Complaint record id # (b)(4).
|