| Catalog Number |
MY8030 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No product or photo was returned by the customer.The customer complaint of leakage could not be verified due to the product not being returned for failure investigation.A device history record review could not be performed on material 24010-0007t because the lot number is unknown.Due to no sample being received, an investigation could not be performed, and a root cause could not be determined.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.Your assistance in this matter has been helpful in trend identification and supporting our commitment to continuous quality improvement.
|
| |
|
Event or Problem Description
|
|
It was reported that bd texium needle-free syringe infusion set was leaking the following information was received by the initial reporter with the following verbatim during doxorubicin infusion, the y-site where the syringe is connected to the primary tubing was leaking when the medication was administered.2 out of 3 syringes leaked at the y-site.May involve my8020/my8030 syringes.
|
| |
|