| Model Number |
406306 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/21/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
During the supraventricular tachycardia procedure, air was aspirated when flushing the inside of the sheath.The sheath was replaced and the procedure was completed with no adverse consequences to the patient.The hospital discarded the reported sheath.Air contamination to the patient has not been confirmed.No additional puncture was performed due to replacing the introducer.
|
| |
|