MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION POLARSHEATH; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION POLARSHEATH; CATHETER, STEERABLE Back to Search Results
Model Number M004CRBS3150
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Low Blood Pressure/ Hypotension (1914); Pericardial Effusion (3271)
Date of Event 11/07/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient became hypotensive and experienced a pericardial effusion, leading to cancellation of the procedure.During a pulmonary vein isolation cryoablation procedure to treat atrial fibrillation, a polar sheath was selected for use.While gaining access and conscious sedation (local anesthesia), the patient's blood pressure dropped, with a systolic reading of 80mmhg.The procedure was continued.Upon the introduction of ice into the heart, the physician performed a baseline assessment of the pericardium and noted no effusions at baseline.A coronary sinus catheter as well as this polarx were introduced.An electrophysiology study was performed prior to transeptal and was non resultant.Transeptal was completed under intracardiac echo with nrg needle and non-boston scientific sheath.The physician successfully crossed, and heparin was administered.The left superior pulmonary vein was wired, and the sheath was exchanged for the polar sheath.When introducing the polar sheath, pressure was needed, and it did 'jump' when crossing the fossa ovalis.The polarxfit balloon was prepped and inserted into the left atrium.During manipulation of the polar map the patient went into atrial fibrillation.The balloon was positioned in the left superior pulmonary vein and ablation was initiated.Isolation of the vein occurred at 47 seconds.The temperature towards the end of the ablation continued to drop until it reached -70 when the physician terminated the ablation at 173 seconds.During this ablation the pressures dropped more to 60mmhg on the readings from the arterial line.At this point the physician checked for effusion on intracardiac echo where there was one apparent around the basal aspect of the left ventricle.At this point the procedure was cancelled, and the physician began a pericardiocentesis.Anesthesia supported the pressure of the patient and heparin reversal was given.Approximately 350cc of blood was removed from the pericardial space.The patient stabilized was extubated and admitted to the hospital.The physician did not note any equipment malfunctions that lead to the effusion.The patient was in stable condition.The device was disposed by the facility and will not be returned as the device performed as expected during the procedure and there were no reports of any performance concerns.
 
Additional Manufacturer Narrative
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARSHEATH
Common Device Name
CATHETER, STEERABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
angela hinkley
4100 hamline avenue north
dc a330
saint paul, MN 55112
6514400979
MDR Report Key20843107
Report Number2124215-2024-74323
Device Sequence Number3917448
Product Code DRA
UDI-Device Identifier00191506039776
UDI-Public00191506039776
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K223824
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/06/2025
Device Model NumberM004CRBS3150
Device Catalogue NumberM004CRBS3150
Device Lot Number0034379694
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2024
Initial Report FDA Received Date12/04/2024
Date Device Manufactured07/06/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-