| Model Number |
5.5X28MM PERIPHERAL SCREW, LOCKING |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 11/05/2024, it was reported by a sales representative via sems-06704164 that an ar-9625 3.0 mm hex driver tip broke off when tightening the locking screw into the mgs baseplate.It was retrieved from the screw head.The surgeon was not using the torque indicator as per their preference when the driver broke, a replacement screwdriver was attached to the torque indicator, and the screw was confirmed to be torqued down.No delay in the case and the case was completed successfully.This was discovered during an rtsa procedure on (b)(6) 2024, with no reported patient harm.Additional information was received on 11/08/2024: the torqued down screw was part number ar-9563-28 lot 15202638.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to over-torquing the screw when inserting into the plate and/or hard bone.
|
| |
|