EXACTECH, INC. LOGIC FEMORAL PS CEM LEFT SZ 3; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-010-01-0230
Medical Device Problem Codes Unstable (1667); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Pain (1994); Synovitis (2094)
Date of Event 08/05/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Report number: 1038671-2023-02159 multiple mdr reports were filed for this event, please see associated report: 1038671-2023-02159.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Event or Problem Description
As reported via legal documentation the patient had a left knee replacement on (b)(6) 2014.Approximately 5 years and 8 months after the initial procedure the patient had a left knee revision on (b)(6) 2020.The patient has experienced mechanical loosening femoral component, damaged poly insert component, synovitis and pain.No further issues or complications were reported.Synovectomy was performed removing scar tissue and inflamed synovium.The tibial insert was noted to be gouged and damaged.The femoral component was grossly loose and separated from the cement.The patient was transported in stable condition to the recovery room.There is no other patient demographic or medical history available.There is no device return.There are no photos or other images of the device provided.No additional information is available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: h6.Mdr section codes updated/corrected: b, c.The revision captured was likely the result of prosthesis wear and an insufficient bond between the femoral component and the bone resulting in aseptic (non-infected) femoral loosening.Possible causes for polyethylene wear include malalignment between the implants, high contact stresses during knee flexion, third body wear, patient-related conditions, instability, or any combination of these possibilities.The extent and root cause of the prosthesis wear and femoral loosening could not be determined as the devices were not returned for evaluation, and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC FEMORAL PS CEM LEFT SZ 3
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
MDR Report Key20844047
Report Number1038671-2024-04617
Device Sequence Number3730231
Product Code JWH
UDI-Device Identifier10885862001146
UDI-Public10885862001146
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K033883
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/24/2019
Device Catalogue Number02-010-01-0230
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2023
Supplement Date Received by Manufacturer03/17/2025
Initial Report FDA Received Date12/04/2024
Supplement Report FDA Received Date03/18/2025
Date Device Manufactured09/26/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age56 YR
Patient SexMale
-
-