| Catalog Number |
9-ASD-012 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a 12mm amplatzer septal occluder was selected for implant on (b)(6) 2024 using a 7f amplatzer trevisio intravascular delivery system.During implant the device took on a cobra shape.The device was not released from the delivery system.The device was removed from the patient and replaced with a new 12mm amplatzer septal occluder.The patient remained hemodynamically stable throughout the procedure.There were no interactions with cardiac structures.There were no angulations or kinks noted on the delivery system.There were no clinically significant delays in the procedure.There were no adverse effects to the patient.The patient status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity was reported.A returned device inspection could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported device deformity could not be conclusively determined.
|
| |
|