| Catalog Number |
121836756 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
11/21/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented in the er with a dislocated hip prosthesis, as the head had escaped the constrained liner.Patient had the large majority of their pelvis removed due to cancer and relies solely on a constrained liner for stability.A new constrained liner and head from another manufacturer, matching the proximal femoral replacement, were replaced uneventfully.No patient harm nor delay is noted.Doi: (b)(6) 2020.Dor: (b)(6) 2024.Affected side: left hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary- no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot- the device lot number is unknown, therefore a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.
|
| |
|
Search Alerts/Recalls
|