| Catalog Number |
HAR17F |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an unknown procedure, during cleaning, the upper part of the jaw, pad, broke.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) date sent: 12/4/2024 d4 batch #: unknown.Additional information was requested and the following was obtained: did the active titanium blade break off? yes did the white tissue pad break off? no is the white tissue pad completely missing from the clamp arm of the device? no did the patient experience any adverse consequences due to this issue? no attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 1/10/2025.D4 batch #: c9dg3k.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the device was returned with the blade tip broken off and not returned.During functional testing on a gen11 an alert screen was displayed.A probable cause of the device stop activating and display an alert screen is blade damage.The device will stop activating when the blade becomes damaged.The device was disassembled to inspect the blade and the blade breaking point was located at an area inside the tube proximal to the tissue pad.This is consistent with a side load being applied to the blade while active with the jaw closed.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result in activation issues such as failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result are such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damage blade can result in a broken blade.No conclusion could be reached as to how this issue occurred.As part of the quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number c9dg3k, and no non-conformances were identified.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 1/9/2025.The device upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
|
| |
|