| Catalog Number |
381437 |
| Medical Device Problem Code |
Retraction Problem (1536)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/03/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device history record review was completed by our quality engineer team for provided material number 381437 and lot number 4037551.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.There are quality controls currently in place to detect this type of defect during the production process.Further action has not been determined necessary at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that bd insyte autoguard needle did not retract.The following information was provided by the initial reporter: a staff member was trying to place a 20g peripheral iv in a patient, but when inserted, the needle would not retract.The white button on the device was stuck and the needle could not be retracted.Therefore, it was removed and another one had to be used resulting in two pokes to the patient.
|
| |
|