• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE, INC. NUVASIVE UNIVERSAL INSTRUMENTS; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE, INC. NUVASIVE UNIVERSAL INSTRUMENTS; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 6904485
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/05/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a procedure, the trial inserter tip fractured off inside the trial.The fractured portion was retrieved along with the trial and no fragment was left in-situ.There was no reported impact to the patient, user, or procedure.No additional information is available.
 
Additional Manufacturer Narrative
Photographs of the device were provided, evaluated, and confirmed the reported event as the inserter tip appeared to be fractured at the bottom thread with the fractured portion completely contained within the associated trial.Operative notes and/or medical records were not provided for review of usage/technique.Based on the information obtained, the root cause of the fracture was likely excessive off-axis and/or twisting motion forces being applied during the attempted insertion into the disc space.Labeling review: ".The implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient." ".Instrument malfunctions which potentially result in surgical delays (thereby causing additional exposure to anesthesia, blood loss, and infection), a change in surgical plan, failure to achieve optimal clinical result and/or, infrequently, cancellation of a procedure.Such malfunctions include instrument fracture, which could make necessary removal difficult or sometimes impossible, with possible consequences of late infection and migration; instrument fracture may also result in injury to the patient." ".Pre-operative warnings: the method of use for the instruments are to be determined by the user¿s experience and training in surgical procedures.The instruments should be treated as any precision instrument and should be carefully placed on trays, cleaned after each use, and stored, according to generally accepted hospital methods and practices.The instruments should be carefully examined prior to use for functionality, excessive wear, or damage.A damaged instrument should not be used as this may increase the risk of malfunction and potential patient injury." ".Intra-operative warnings: the physical characteristics required for many instruments do not permit them to be manufactured from implantable materials.If any broken fragments of instruments remain in the body of a patient, they could cause allergic reactions or infections.If an instrument breaks in surgery and fragments go into the patient, these pieces should be removed prior to closure and should not be implanted.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE UNIVERSAL INSTRUMENTS
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
NUVASIVE, INC.
7475 lusk boulevard
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INC.
7475 lusk boulevard
san diego CA 92121
Manufacturer Contact
macy stout
7475 lusk boulevard
san diego, CA 92121
MDR Report Key20848720
Report Number2031966-2024-00323
Device Sequence Number3930157
Product Code LXH
UDI-Device Identifier00887517389091
UDI-Public887517389091
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6904485
Device Lot NumberEM41066
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/05/2024
Initial Report FDA Received Date12/04/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/18/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ASSOCIATED TRIAL
Patient Age68 YR
Patient SexFemale
-
-