| Model Number |
3030007 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received from user facility via a manufacturer representative regarding a device used for spinal therapy.It was rep orted that, the drivers were stripped.There was no patient involvement reported.No further complications reported regarding the event.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis of part # 3030007 ; lot # im18f014 visual and optical examination revealed the t8 portion of the driver¿s tip has broken and the t10 tip is stripped.The edges have been rolled over and deformed.This type of damage is consistent with torsional overload.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|