• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONMED UTICA CONMED; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONMED UTICA CONMED; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 138110
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Received 2 items 138110 in their unopened original packaging.Performed a visual inspection of the devices, there were signs of a breach in both devices.No need for a functional test since there were breaches in the heat seals.Root cause cannot be determined, however, based upon evaluation of the device and ifu; a possible cause of this event could be that the packages were improperly sealed.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.A two-year lot history review shows a total of 2 devices for this lot number and failure mode.A two-year review of complaint history revealed there has been a total of 207 complaints, regarding (b)(4), for this device family and failure mode.During this same time frame (b)(4) have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use the user is advised that sterility guaranteed unless package has been opened, broken or damaged.Inspect and test each device before use.We will continue to monitor per regulatory requirements to help ensure patient safety.
 
Event or Problem Description
During incoming inspection, the distributor rejected this device, 138110, for an insufficient heatseal.There was no contact with the patient as this was found during incoming inspection.This will be reported as a malfunction with potential for injury upon reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONMED
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer (Section G)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer Contact
beth madeline
11311 concept blvd
largo, FL 33773
7274708358
MDR Report Key20852568
Report Number1320894-2024-00417
Device Sequence Number8039640
Product Code GEI
UDI-Device Identifier30653405003452
UDI-Public(01)30653405003452(17)250927(10)202009281
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K943542
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 12/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/27/2025
Device Catalogue Number138110
Device Lot Number202009281
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2024
Initial Report FDA Received Date12/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/28/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-