| Model Number |
GL358XL/LC358XL |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Provider alleges the consumer alleges the hand control caught on fire and melted the cords.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not yet been made available for evaluation.Should further information or the device become available, a follow-up report will then be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
The root cause of the failure appears to be a damaged strain relief causing harness failure; the pcb was unaffected.The product literature contains instruction to periodically check the hand control for visible damage.
|
| |
|
Event or Problem Description
|
|
Provider alleges the consumer alleges the hand control caught on fire and melted the cords.
|
| |
|
Search Alerts/Recalls
|