| Model Number |
SUTRLASSO SD SFT,45°RT CRV,R |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/31/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 11/06/2024, it was reported by a sales representative via (b)(4) that (qty.2) of an ar-5068-45r suturelasso¿ sd shaft were broken off while attempting to pierce the labrum.All broken pieces were retrieved from the patient and the case was completed successfully.This was discovered during a shoulder labral repair procedure (b)(6) 2024, with no reported patient harm.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force used during suture passing/tightening /tensioning.
|
| |
|