• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® I.C. ALL-SILICONE TEMP SENSING FOLEY CATHETER; TEMPERATURE SENSING FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® I.C. ALL-SILICONE TEMP SENSING FOLEY CATHETER; TEMPERATURE SENSING FOLEY CATHETER Back to Search Results
Catalog Number 119310M
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Codes Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/12/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported by the critical care nurses in an online survey that the patient experienced an adverse event directly caused or attributable to the usage of device and resulted in the following complications bypassing.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Correction: b, f, h.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the critical care nurses reported in an online survey that the patient experienced and adverse event directly caused or attributable to the usage of device and the resulted in the following complications bypassing.Per additional information received via mail on 13dec2024, stated that the respondent indicated the patient did required medical or surgical intervention and indicated surgery.
 
Event or Problem Description
It was reported that the critical care nurses reported in an online survey that the patient experienced and adverse event directly caused or attributable to the usage of device and the resulted in the following complications bypassing.Per additional information received via mail on 13dec2024, stated that the respondent indicated the patient did required medical or surgical intervention and indicated surgery.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was available for evaluation and further investigation did not result in any additional findings.No actions can be taken at this time since a root cause was not identified.The dhr review could not be performed without a lot number.The instructions for use were found adequate and state the following: caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Do not aspirate urine through drainage funnel wall.Aggressive traction, particularly in the presence of suturing, is not recommended for 100 percent silicone balloon foley catheters.As with all temperature probes: in the presence of rf energy sources, local heating, temperature errors, and probe damage may occur.In medical use, unplug the temperature-sensing catheter at the extension cable before activating electrosurgical or other types of direct coupled rf energy sources.Warning: this product should never be connected to the temperature monitor or connected to a cablle during an mri procedure.Failure to follow this guidelline may resullt in serious injury to the patient.Refer to the instructions for use.It is important to closely follow these specific conditions that have been determined to permit the examination to be conducted safely.Any deviation may result in a serious injury to the patient.Valve type: use luer slip syringe.Do not use needle.Do not stretch catheter.This will cause repositioning of probe.Do not use stylet.This will cause stretching of catheter.Recommended inflation capacities 14 fr.And larger (5cc balloon): use 10cc sterile water do not exceed recommended capacities.Visually inspect the product for any imperfections or surface deterioration prior to use.Catheters should be replaced in accordance with the cdc guideline guideline for prevention of catheter-associated urinary tract infection.At the onset or first signs of a urinary tract infection, catheter encrustation, or any other catheter-related adverse effect, the catheter should be replaced.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe stick in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRI-SIL® I.C. ALL-SILICONE TEMP SENSING FOLEY CATHETER
Common Device Name
TEMPERATURE SENSING FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key20853533
Report Number1018233-2024-07781
Device Sequence Number17086495
Product Code MJC
UDI-Device Identifier00801741039881
UDI-Public(01)00801741039881
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K070582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number119310M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/12/2024
Supplement Date Received by Manufacturer12/13/2024
01/10/2025
Initial Report FDA Received Date12/05/2024
Supplement Report FDA Received Date01/07/2025
01/23/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexUnknown
-
-