SMITH & NEPHEW, INC. GII OXINIUM FEM SIZE 7 LEFT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
71420140 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
11/12/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, after primary tka surgery had been performed on (b)(6) 2012, the patient started developing anterior knee pain.This adverse event was treated by resurfacing the patella on (b)(6) 2024.There was no explantation of one (1) journ art ins bcs std 3-4 lt 9, one (1) gii oxonium fem size 7 left, one (1) gii c/r art ins sz 7-8 9mm, one (1) gns ii cmt tib size 7 left and one (1) journ art ins bcs std 3-4 rt 9 since just the patella was resurfaced.Patient's current health status is unknown.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3,h6: this complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Search Alerts/Recalls
|
|
|