• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMBU A/S VIVASIGHT 2 DLT KIT 39 FR - LEFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMBU A/S VIVASIGHT 2 DLT KIT 39 FR - LEFT Back to Search Results
Model Number 412391000
Medical Device Problem Code Structural Problem (2506)
Health Effect - Clinical Code Upper Respiratory Tract Injury (4599)
Date of Event 11/06/2024
Type of Reportable Event Serious Injury
Event or Problem Description
The incident leading to tonsil injuries occurred in two separate procedures (the second incident is covered by 1220828-2024- 00024).A crna was trying to intubate with the vivasight 2 dlt product and they had to adjust the distance of tip of the tube past the vocal cords.The curve at the end of the bronchial lumen was more angulated than a typical robert-shaw double lumen endotracheal tube, and got more easily caught on the tracheal rings.One could either advance the bronchial cuff further past the vocal cords with the stylet, or if the stylet was removed after the tip was in the trachea, one could turn the tube 120-180 degrees leftward while advancing and then return it to 90 degrees once past the ring that the tip got hung up on.
 
Additional Manufacturer Narrative
Measurement from end to tip of endotracheal bevel was performed, in order to evaluate if the bending is within specifications (15°-53°).The angle of the retention sample was 60°, thus outside of specifications.Corrective action had been initiated to investigate this issue and the preliminary investigation showed that the bending was caused by mishandling by manufacturing personnel, in which additional bending of endobronchial tip occurred after stylet insertion process.There is an ongoing recall 'ambu2024fa00001' which was initiated due to hyperangulation of the distal tip.The lot number of the affected product involved in this incident corresponds to the lot numbers affected by the recall.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVASIGHT 2 DLT KIT 39 FR - LEFT
Common Device Name
VIVASIGHT 2 DLT KIT 39 FR - LEFT
Manufacturer (Section D)
AMBU A/S
baltorpbakken 13
ballerup, denmark 2750
DA  2750
MDR Report Key20854058
Report Number1220828-2024-00025
Device Sequence Number12971493
Product Code CBI
UDI-Device Identifier05707480145768
UDI-Public05707480145768
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/05/2024
Report Date (Section F) 11/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number412391000
Device Catalogue Number412391000
Device Lot Number1000989835
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date11/06/2024
Event Location Hospital
Date Report to Manufacturer11/06/2024
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date12/05/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-