• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. SFXSPI PDS+ BI VIO 14IN2 2-0 DA CT-1; STRATAFIX SPIRAL PDS PLUS BIDIRECTIONAL KNOTLESS TISSUE CONTROL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. SFXSPI PDS+ BI VIO 14IN2 2-0 DA CT-1; STRATAFIX SPIRAL PDS PLUS BIDIRECTIONAL KNOTLESS TISSUE CONTROL DEVICE Back to Search Results
Catalog Number SXPP2B412
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/08/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to the sales rep that during an unknown orthopedic procedure that while suturing the subcuticular layer of tissue two sutures broke around the middle of the suture.Another like suture was used to continue with the procedure.There were no adverse consequences for the patient.Two sutures returning.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 12/5/2024.H6 component code: g07002 - no device problem found.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to ethicon for evaluation.One needle suture in two sections and a top foil were received for analysis.The product code sxpp2b412 is double-armed.During visual inspection of the sample, the swage and attachment area were noted as expected in the needles.However, marks by use were observed on the body needles.The ends of the suture pieces were split probably by a surgical instrument.Also, the ends match with each other.The product code sxpp2b412 contains an absorbable suture.Since the sample was received open, the functional test could not be performed due to undetermined exposure time to the environment.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.No conclusion could be reached as to what caused the reported complaint.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SFXSPI PDS+ BI VIO 14IN2 2-0 DA CT-1
Common Device Name
STRATAFIX SPIRAL PDS PLUS BIDIRECTIONAL KNOTLESS TISSUE CONTROL DEVICE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km 8.3
san lorenzo PR 00754
*   00754
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key20854217
Report Number2210968-2024-12879
Device Sequence Number18350723
Product Code NEW
UDI-Device Identifier10705031464520
UDI-Public(01)10705031464520(17)251130(10)TPBEKD
Combination Product (Y/N)Y
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K192144
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2025
Device Catalogue NumberSXPP2B412
Device Lot NumberTPBEKD
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2024
Initial Report FDA Received Date12/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-