• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE 3-LUMEN SPHINCTEROTOME V; SINGLE USE 2-LUMEN SPHINCTEROTOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE 3-LUMEN SPHINCTEROTOME V; SINGLE USE 2-LUMEN SPHINCTEROTOME Back to Search Results
Model Number KD-V411M-0730
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
It was reported that while using the single use 3-lumen sphincterotome v the cutting wire of the papillotome broke or was damaged when cutting under tension.The reported malfunction occurred during a therapeutic endoscopic retrograde cholangio-pancreatography.The procedure was completed using a similar device.There were no reports of patient harm.
 
Additional Manufacturer Narrative
Corrected fields: d7a.Updated fields: h2, d4 - serial #, d4 - lot #, h3, h6, h11.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: the coated portion of the cutting wire was torn a definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the cutting wire was broken at proximal end site.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
No additional information received from the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE 3-LUMEN SPHINCTEROTOME V
Common Device Name
SINGLE USE 2-LUMEN SPHINCTEROTOME
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA   036-0357
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20854239
Report Number9614641-2024-20052
Device Sequence Number8831966
Product Code KNS
UDI-Device Identifier04953170184055
UDI-Public04953170184055
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K950166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberKD-V411M-0730
Device Lot Number3XV 17
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/06/2024
Supplement Date Received by Manufacturer01/10/2025
Initial Report FDA Received Date12/05/2024
Supplement Report FDA Received Date02/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-