| Model Number |
KD-V411M-0730 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that while using the single use 3-lumen sphincterotome v the cutting wire of the papillotome broke or was damaged when cutting under tension.The reported malfunction occurred during a therapeutic endoscopic retrograde cholangio-pancreatography.The procedure was completed using a similar device.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected fields: d7a.Updated fields: h2, d4 - serial #, d4 - lot #, h3, h6, h11.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: the coated portion of the cutting wire was torn a definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the cutting wire was broken at proximal end site.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|