| Model Number |
N/A |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the liner was unable to seat.A new liner was required to be open and seated immediately.No known impact or consequence to the patient.Attempts have been made, and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: unknown cup unknown.G2: foreign: australia.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the liner was unable to seat.A new liner was required to be open and seated immediately.The initial cup implant was implanted into the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.Updated: a1, a2, a3, b5, d10, g3, g6, h2.D10: 010000704 g7 bonemaster ltd acet shl 54f 7772360.
|
| |
|
Event or Problem Description
|
|
No additional event information to report at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.Visual examination of the returned product identified indentations to the outside radius of the liner.The locking feature of the device had been damaged and there were visible shavings.The height and diameter were measured and found to be within specification.Review of the device history records identified no deviations or anomalies during manufacturing.Review of complaint history found no additional related issues for this item and the reported part and lot combination.Medical records were not provided.A definitive root cause cannot be determined.The complaint cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|