MAUDE Adverse Event Report: ARTHREX, INC. UNIVERS APEX HUMERAL STEM, 11MM; PROSTHESIS, SHOULDER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. UNIVERS APEX HUMERAL STEM, 11MM; PROSTHESIS, SHOULDER, CEMENTED Back to Search Results
Model Number UNIVERS APEX HUMERAL STEM, 11MM
Medical Device Problem Codes Material Fragmentation (1261); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Implant Pain (4561)
Date of Event 11/04/2024
Type of Reportable Event Serious Injury
Event or Problem Description
On 11/07/2024, it was reported by a sales representative via phone that a patient had a shoulder scope arthroplasty procedure completed successfully three years ago.The patient reported having post-surgery adverse reactions to the implant that was implanted.The surgeon performed a biopsy and found metallosis in the joint.Photos were taken of the issue.Additional information was received on 11/08/2024 via email: on (b)(6) 2024, the surgeon brought one of his patients back to the operating room for a biopsy and diagnostic arthroplasty to see why, after a successful tsa outcome, the patient had suddenly experienced a high amount of pain in the last 6 months.Shots and conservative treatments failed, and the patient was brought into the operating room for arthroscopic evaluation.The surgeon found severe metallosis in the joint.No failure of implants or looseness could not be determined at the time.The surgeon took the arthroscopic biological cultures and sent them to the lab for analysis.The arthrex products implanted in the initial surgery were ar-9106-03 arthrex univers vaultlock glenoid, ar-9100-11s univers apex humeral stem, and ar-9156-22p arthrex univers ii 3d humeral head.The initial surgery was done on (b)(6) 2021 at the same location as the diagnostic scope.Additional information has been requested.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
Additional information was received on 11/22/2024 via email: the reaction was localized in the shoulder joint, and no revision has been scheduled yet.
 
Additional Manufacturer Narrative
Additional information: d6a, g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to a patient-specific event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERS APEX HUMERAL STEM, 11MM
Common Device Name
PROSTHESIS, SHOULDER, CEMENTED
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key20855660
Report Number1220246-2024-08710
Device Sequence Number18336917
Product Code KWS
UDI-Device Identifier00888867134652
UDI-Public(01)00888867134652(11)170131(17)211231(10)10083346
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K131633
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2021
Device Model NumberUNIVERS APEX HUMERAL STEM, 11MM
Device Catalogue NumberAR-9100-11S
Device Lot Number10083346
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2024
Supplement Date Received by Manufacturer11/22/2024
06/26/2025
Initial Report FDA Received Date12/05/2024
Supplement Report FDA Received Date12/20/2024
07/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/31/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-