EXACTECH, INC. OPTETRAK ASY,PS CEMENTED FEMORAL, SZ 4,; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
234-03-04 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Scar Tissue (2060); Osteolysis (2377); Joint Laxity (4526); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
05/18/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
1038671-2023-02675 d10: concomitants: aa1350 1510-s - cemex system fast 70 gm 1580510 200-02-38 - three peg patella 38mm 1578514 200-04-44 - cemented finned tib.Tra sz 4f/4t 98033 203-96-40 - (11-3973) sys 6 90x25x1.27 85104 203-96-42 - 11-4132 sys 6 90x13/21x1.19 1347522 234-03-04 - optetrak asy,ps cemented femoral, sz 4, aa1237 asa0030 - sterile disposable containers multiple mdr reports were filed for this event, please see associated reports: 1038671-2023-02675 the reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, tibial loosening, and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
As reported via legal documentation, a patient had right knee replacement.They subsequently underwent right knee revision surgery, approximately 162 months post primary procedure.Failed right total knee arthroplasty due to mechanical loosening and catastrophic polyethylene wear resulting in severe osteolysis, scar tissue and joint laxity.The patient tolerated the procedure well, was awoken and was brought to the recovery room in stable condition.There is no other patient demographic or medical history available.There is no device return.There are no photos or other images of the device provided.No additional information is available.
|
| |
|
Search Alerts/Recalls
|
|
|