• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION INTRAVIA; CONTAINER, I.V.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION INTRAVIA; CONTAINER, I.V. Back to Search Results
Model Number 2J8003
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/08/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that a clear piece of plastic was found inside the solution/fluid path of an unspecified quantity of 500ml non-dehp fluid path intravia containers.This was observed prior to spiking or reconstitution.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
Additional information: d9, h3, h4, h6, h11.H11: the actual device and two companion samples were received for evaluation.A visual inspection was performed on the actual sample, and it was noted that a small particulate was inside of the actual sample (a small clear plastic piece floating in the bag).Visual inspection of the companion samples did not identify any abnormalities that could have contributed to the reported condition.The reported condition was verified for the actual sample.The cause of the reported condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRAVIA
Common Device Name
CONTAINER, I.V.
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
piisa industrial park antigua
carretera sanchez km 18 1/2
haina, san cristobal 91000
DR   91000
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key20857037
Report Number1416980-2024-07051
Device Sequence Number18325387
Product Code KPE
UDI-Device Identifier00085412029566
UDI-Public(01)00085412029566(10)DR24F06031
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K964853
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number2J8003
Device Catalogue Number2J8003
Device Lot NumberDR24F06031
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2024
Supplement Date Received by Manufacturer01/02/2025
Initial Report FDA Received Date12/05/2024
Supplement Report FDA Received Date01/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/07/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NA.
Patient SexUnknown
-
-