• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. LGC TIBIAL FIT TRAY CEM SZ 3F / 3T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LGC TIBIAL FIT TRAY CEM SZ 3F / 3T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-012-45-3030
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Scar Tissue (2060); Osteolysis (2377); Implant Pain (4561); Swelling/ Edema (4577)
Date of Event 04/12/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Report number: 1038671-2023-02944, 1038671-2024-04643 d10 concomitants: 02-010-03-0330 - logic cr femoral cem, right, sz 3 4383932 02-012-45-3030 - lgc tibial fit tray cem sz 3f / 3t 4184941 02-012-48-3013 - logic cr tib insert slope+, sz 3, 13mm 2235380 13a2101 - cemex system fast genta 70g aa7042 200-05-26 - inset patella 26mm 3971230 asa0030 - sterile disposable containers aa9222 multiple mdr reports were filed for this event, please see associated reports: 1038671-2023-02944, 1038671-2024-04643 the reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, tibial loosening, and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Event or Problem Description
As reported via legal documentation the patient had a right knee replacement on (b)(6) 2017.Approximately 5 years and 2 months after the initial procedure the patient had a right knee revision on (b)(6) 2022.The previous liner was dislodged from the tibial component.Upon inspection there was severe delamination wear on the surface of the polyethylene.Both tibial and femoral components had area of severe osteolysis leading to debonding at the bone cement joint.Patellar component found to be significantly worn.The patient tolerated the procedure well.There is no other patient demographic or medical history available.There is no device return.There are no photos or other images of the device provided.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGC TIBIAL FIT TRAY CEM SZ 3F / 3T
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
MDR Report Key20858075
Report Number1038671-2024-04645
Device Sequence Number3917824
Product Code JWH
UDI-Device Identifier10885862003966
UDI-Public10885862003966
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K101981
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/18/2025
Device Catalogue Number02-012-45-3030
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/25/2024
Initial Report FDA Received Date12/05/2024
Date Device Manufactured11/19/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age69 YR
Patient SexMale
-
-