| Model Number |
CV-1214 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/27/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned.Three retained samples from the same lot were used to perform a knot pull tensile test.The retained samples used for the knot pull tensile test met requirements.Production records were reviewed.There were no nonconformities noted with the lot during production.All finished goods testing requirements were met prior to release.There was no evidence that the device failed to meet specifications, and the report could not be substantiated.A cause for the event cannot be established.This report and use of categorical definitions required by fda 3500a does not constitute an admission by riverpoint medical or its employees that riverpoint medical or its employees have caused or contributed to the event described in this report.Riverpoint medical has filed this information to comply with the medical device reporting regulation 21 cfr 803.If additional information is provided to riverpoint medical regarding this event, a supplementary 3500a form will be submitted as required by the fda.
|
| |
|
Event or Problem Description
|
|
According to the reporter, "it was reported that during a procedure involving the use of velosorb* 4-0 sutures, the suture line did not hold.The same issue occurred with the second suture.".
|
| |
|
Search Alerts/Recalls
|